Comparison of artifact from craniomaxillofacial internal fixation devices: magnetic resonance imaging
Abstract:
This study compares artifact from craniomaxillofacial internal fixation devices in magnetic resonance images and examines heating and magnetic deflection effects on these devices. Stainless steel wires, microfixation plates of vitallium and titanium, and minifixation and mandibular reconstruction plates of stainless steel, vitallium, and titanium were evaluated. The plates were mounted on a dry skull and submerged in dilute copper sulfate solution. All images were obtained in duplicate by two independent, nonblinded teams of observers with a 1.5-T Signa magnetic resonance system. Each team ranked artifact size for each material by computerized measurement. The rank order of artifact size for each material within each fixation system group and between fixation groups was identical between the two teams. Bath and plate temperatures were recorded before and after imaging. Artifact production was related to hardware size and composition. Titanium hardware caused the least amount of "black-hole" artifact. Vitallium and stainless steel fixation devices of similar size produced significantly more artifact. No significant heating or magnetic deflection effects were seen with any of the fixation devices.
More Related Videos
07:33In Vivo Evaluation of Fracture Callus Development During Bone Healing in Mice Using an MRI-compatible Osteosynthesis Device for the Mouse Femur
Published on: November 14, 2017
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI