Related Experiment Videos
[Planimetric quantification of aortic valve stenoses using multiplanar transesophageal echocardiography]
1Medizinische Klinik III-Kardiologie, Herzzentrum Kaiser-Wilhelm-Krankenhaus, Duisburg.
Abstract:
Standard techniques used in order to quantify the severity of aortic valve stenoses in clinical practice comprise: transthoracic echocardiography, namely, by determining maximum and mean transvalvular gradients and evaluating aortic valve areas, as well as invasive techniques which quantify aortic valve areas through hemodynamic pressure measurements and application of the Gorlin formula. Since the introduction of the multiplane TEE technique, it has become feasible to scan the aortic valve in a strictly horizontal plane and quantify the aortic valve orifice by planimetry. In this study, we investigated 23 patients with various degrees of aortic valve stenoses. We compared aortic valve areas, which had been planimetrically determined by multiplane TEE scans, and mean aortic valve gradients (standard TEE technique) with pressure gradients and valve areas derived from hemodynamic measurements obtained during cardiac catheterization, and have found that the valve areas as well as the mean pressure gradients correlate well.