Artifacts in fetal pulse oximetry: incomplete sensor-to-skin contact
J O Gardosi1, D Damianou, C M Schram
1Department of Obstetrics and Gynaecology, University of Nottingham, United Kingdom.
Objective:
Our purpose was to investigate apposition of reflectance sensors as a possible source of artifact in fetal pulse oximetry.
Study Design:
A laboratory model was created to examine the effect of varying the distance of a reflectance probe from a subject's finger. The setup used two wavelengths (660 and 940 nm) and pulsatile (AC) and nonpulsatile signals (DC). An investigator's finger was stabilized in a purposely built mold from where the distance between skin and sensors could be varied in 1 mm steps.
Results:
The AC and DC components of the signal react differently to increasing the distance between sensor and skin. The overall effect is that the ratio [formula: see text] increases with skin to sensor distance, which results in falsely low readings of oxygen saturation.
Conclusions:
Poor sensor-to-skin contact in reflectance pulse oximetry may cause false readings despite the presence of good waveforms and hence go unrecognized. Probe and software design needs to be able to identify and exclude such artifact before this technique can become useful for intrapartum fetal monitoring.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:44Near-Infrared Spectroscopy During Reactive Hyperemia for the Assessment of Lower Limb Vascular Function
Published on: March 22, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring pulse
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse
The pulse serves as a clinical indicator...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...