Intraoperative cardiac output monitoring: comparison of impedance cardiography and thermodilution
Related Experiment Video
Updated: Oct 20, 2026

Impedance Pneumography for Minimally Invasive Measurement of Heart Rate in Late Stage Invertebrates
Published on: April 4, 2020
A C Perrino1, A Lippman, C Ariyan
1Department of Anesthesiology, Yale University School of Medicine, New Haven, CT 06510.
Abstract:
Impedance cardiography (IC) is a noninvasive, simple to use method of cardiac output (CO) determination. A prospective evaluation of IC monitoring was performed in 50 patients undergoing noncardiac surgery. IC CO measurements (NC-COM3-Revision 7, BoMed Manufacturing) were compared to simultaneous measurements of thermodilution (TD) CO to assess the validity of this technique for intraoperative cardiac monitoring. Adequate impedance signals could not be obtained in 7 of the 50 patients. IC CO measurements were highly correlated to TD CO (P < .005), with a correlation coefficient r = 0.84. Bias analysis, however, indicated clinically significant disagreement between the two techniques. IC CO tended to underestimate TD CO (mean bias = -0.41 L/min) and the SD of the bias was 1.0 L/min (95% level of agreement 1.6 to -2.4 L/min). Trending data showed IC to accurately track the direction of TD CO changes but to underestimate their magnitude (r = 0.60, intercept -0.7 L/min, slope 0.47). Factors that may have impaired the performance of IC in this study include the high prevalence of cardiac disease in the study population and electrical noise in the operative setting. Further development of IC appears warranted if it is to prove useful as an intraoperative cardiac monitor.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...