A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Head position and oral vs nasal route as factors determining endotracheal tube resistance
1Department of Anesthesiology, University of Washington School of Medicine, Seattle.
Study Objective:
We performed this in vitro study to determine the following: (1) if there is any significant difference in resistance between comparably sized endotracheal tubes (ETTs) in simulated anatomic oral and nasal conformations: (2) if neck flexion would increase the resistance of the ETT; and (3) if a wire-reinforced tube in simulated oral conformation would minimize the resistance increase at bends in the tube.
Design:
The pressure drops (the change in pressure the flow through the ETT) at the proximal end of three sizes of tubes (6-, 7-, and 8-mm inner diameter) were measured in anatomic conformations at flows ranging from 20 to 100 L/min with the tubes warmed to 37 degrees C.
Results:
There were no significant differences in pressure drops between comparably sized ETTs in the nasal vs oral conformation at any flow tested. Maximal head flexion produced a small increase in pressure drop for the standard 6-mm ETT but not for the larger tubes. The wire-reinforced tubes, when compared with the standard 7- and 8-mm ETTS, actually had greater pressure drops across the tubes at high flows.
Conclusions:
(1) Pressure differences between ETTs in nasal and oral conformations with comparable size and length are insignificant, even at high flow rates. (2) Maximal neck flexion does not deform the tube enough to make a clinically significant difference in resistance in vitro, although the 6-mm ETT had a slightly higher resistance in flexion. (3) A size 7- or 8-mm reinforced ETT has a higher resistance compared with a standard ETT.
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques