Reproducibility of postural changes in intraocular pressure with the Tono-Pen and Pulsair tonometers
M R Wilson1, R S Baker, P Mohammadi
1Charles R. Drew University of Medicine and Science, Los Angeles, California.
Abstract:
Reproducibility of measurements of postural changes in intraocular pressure was determined by using the Tono-Pen (Mentor O & O, Inc., Norwell, Massachusetts) and Pulsair (Keeler Instruments, Inc., Broomhall, Pennsylvania) tonometers. Thirty subjects had three repeated sitting and reclining measurements performed on three separate visit days. Reproducibility coefficients of 3.3% for the Tono-Pen and 6.3% for the Pulsair were obtained for the within-visit postural change measurements. Across-visit reproducibility coefficients were 7.9% and 26.2% for the Tono-Pen and Pulsair, respectively. The estimated standard deviations for both the within-visit and across-visit postural change measurements were high for both instruments. These results indicate poor reproducibility of measurements of postural changes in intraocular pressure.