Quantitative analysis of four EMG amplifiers

E J Perreault1, I W Hunter, R E Kearney

  • 1Biomedical Engineering Department, McGill University, Montreal, Québec, Canada.

Journal of Biomedical Engineering
|September 1, 1993
PubMed
Summary

Testing four electromyography (EMG) amplifiers revealed varied performance. Proper skin preparation is crucial, as impedance mismatches significantly reduce common mode rejection ratio (CMRR) and noise rejection in EMG equipment.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Online adaptive neural control of a robotic lower limb prosthesis.

Journal of neural engineering·2017
Same author

Response to Hughes et al., critique of our study on the effects of large-scale closures in the GBR.

Ecological applications : a publication of the Ecological Society of America·2016
Same author

Jet injection of a monoclonal antibody: A preliminary study.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference·2016
Same author

Large-scale expansion of no-take closures within the Great Barrier Reef has not enhanced fishery production.

Ecological applications : a publication of the Ecological Society of America·2015
Same author

Automated analysis of respiratory behavior in extremely preterm infants and extubation readiness.

Pediatric pulmonology·2015
Same author

A new movement artifact detector for photoplethysmographic signals.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference·2013

Area of Science:

  • Biomedical Engineering
  • Neuroscience
  • Signal Processing

Background:

  • Electromyography (EMG) amplifiers are essential for measuring muscle electrical activity.
  • Variability in EMG amplifier performance can impact data quality.
  • Understanding equipment characteristics is key for accurate physiological measurements.

Purpose of the Study:

  • To quantitatively assess the diversity and specificity of four typical EMG amplifiers.
  • To characterize key performance metrics including gain, phase, common mode rejection ratio (CMRR), and noise.
  • To evaluate the impact of electrode-skin impedance on amplifier performance.

Main Methods:

  • Quantitative testing of four EMG amplifier models.
  • Measurement of gain, phase response, CMRR, and noise characteristics.
  • Introduction of controlled input impedance mismatches to simulate real-world conditions.

Main Results:

  • Observed significant diversity in gain and phase responses among amplifiers, suggesting suitability for different applications.
  • Demonstrated a dramatic decrease in CMRR for all amplifiers when input impedance mismatches exceeded 10 kΩ.
  • Highlighted the critical role of electrode-skin impedance in determining amplifier noise rejection.

Conclusions:

  • Proper skin preparation is essential for optimizing the performance of EMG amplifiers.
  • Impedance mismatches commonly encountered in surface EMG can severely compromise signal quality.
  • Careful consideration of amplifier specifications and preparation protocols is necessary for reliable EMG data acquisition.