Temporal lobes: surface MR coil phased-array imaging
Related Experiment Video
Updated: Nov 8, 2026

Three-Dimensional Phase Resolved Functional Lung Magnetic Resonance Imaging
Published on: June 21, 2024
C E Hayes1, J S Tsuruda, C M Mathis
1Department of Radiology, University of Washington, Seattle 98195.
Abstract:
The authors describe a four-coil surface phased-array coil for magnetic resonance (MR) imaging of the temporal lobes. On the basis of measurements in a phantom, the signal-to-noise ratio of the array is 1.67 times or more better than that obtained with a standard quadrature head coil in the region of the hippocampi. The higher sensitivity of this phased array permits use of imaging pulse sequences for which the head coil is inadequate.