Related Experiment Video
Updated: Aug 7, 2026

Safety Precautions and Operating Procedures in an (A)BSL-4 Laboratory: 3. Aerobiology
Published on: October 3, 2016
Microbiological hazard from the exhaust of a high-vacuum sterilizer
Abstract:
Data are presented which show the potential for release of viable microorganisms into the atmosphere from high-vacuum steam sterilizers during the evacuation cycle preceding application of steam under pressure. Bacillus subtilis var. niger spores, Serratia marcescens cells, and T1 coliphage disseminated into the sterilizer chamber as small particles from liquid suspensions, and dried spores of B. subtilis var. niger distributed on bulk discard materials were recovered from the atmosphere around pipes venting steam from the steam ejectors used to create chamber vacuum. Evaluation of the hazard involved is discussed, and the design, fabrication, and installation of a valved filter system for preventing release of viable microorganisms are presented. The filtration system utilized an F-700 water-resistant filter and was shown to eliminate the release of viable airborne microorganisms from a high-vacuum sterilizer. A method is presented for determining size requirements for an atmospheric vent filter in relation to the volume of a sterilizer.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Key Techniques in Microbiology
Scale-Up Processes