Cluster of Enterobacter cloacae pseudobacteremias associated with use of an agar slant blood culturing system

M L Pearson1, D A Pegues, L A Carson

  • 1National Center for Infectious Diseases, Centers for Disease Control and Prevention, Atlanta, Georgia 30333.

Insights

A new blood culturing system at Shiprock Hospital led to an Enterobacter cloacae outbreak. Breaks in aseptic technique and lab contamination caused pseudoinfections, highlighting the need for strict protocols.

Area of Science:

  • Clinical Microbiology
  • Infectious Diseases
  • Laboratory Science

Background:

  • An unusual increase in Enterobacter cloacae (E. cloacae) positive blood cultures was observed at Shiprock Hospital.
  • The rise coincided with the implementation of a new blood culturing system.

Purpose of the Study:

  • To investigate the cause of the E. cloacae outbreak.
  • To identify risk factors associated with contaminated blood cultures.

Main Methods:

  • A case-control study was conducted comparing E. cloacae positive cultures with negative cultures.
  • Analysis focused on suspected pseudoinfections (cultures from patients without septicemia symptoms).
  • Investigated aseptic technique, equipment integrity, and environmental contamination, including plasmid analysis of isolates.

Main Results:

  • Blood samples processed during midnight shifts and incubated with other positive samples showed increased contamination risk.
  • Breaks in aseptic technique and leakage from the new culturing system were identified.
  • E. cloacae isolates from patients, technicians, and the environment shared identical plasmid profiles, indicating a common source.

Conclusions:

  • The outbreak was attributed to environmental contamination and breaches in aseptic technique linked to the new blood culturing system.
  • Emphasizes the critical importance of rigorous aseptic practices, quality control, and environmental monitoring in clinical laboratories, especially during system changes.