Portable data acquisition and control apparatus for implanted drug infusion pump interrogation

A S Cress1, B D Wigness, F D Dorman

  • 1Division of Biomedical Engineering, University of Minnesota, Minneapolis 55455.

ASAIO Journal (American Society for Artificial Internal Organs : 1992)
|July 1, 1993
PubMed
Summary

Diagnosing implantable drug infusion pump failures is crucial. A new portable device monitors pump performance by measuring reservoir pressure during refills, simplifying failure detection.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

IFSO Worldwide Survey 2016: Primary, Endoluminal, and Revisional Procedures.

Obesity surgery·2018
Same author

Erratum to: Bariatric Surgery and Endoluminal Procedures: IFSO Worldwide Survey 2014.

Obesity surgery·2017
Same author

Bariatric Surgery and Endoluminal Procedures: IFSO Worldwide Survey 2014.

Obesity surgery·2017
Same author

Reply to Letter to the Editor: Bariatric Surgery Worldwide 2013 Reveals a Rise in Mini-Gastric Bypass.

Obesity surgery·2015
Same author

Bariatric Surgery Worldwide 2013.

Obesity surgery·2015
Same author

Systematic review and meta-analysis of medium-term outcomes after banded Roux-en-Y gastric bypass.

Obesity surgery·2014

Area of Science:

  • Biomedical Engineering
  • Medical Device Technology
  • Drug Delivery Systems

Background:

  • Implantable drug infusion pumps are widely used clinically.
  • Methods for diagnosing suspected pump failures are needed.
  • Previous benchtop apparatus was cumbersome for in vivo monitoring.

Purpose of the Study:

  • To develop a portable apparatus for diagnosing implantable drug infusion pump failures.
  • To provide a simpler and more accessible method for in vivo pump monitoring.

Main Methods:

  • A portable interrogation unit was created, comprising a laptop computer, data acquisition software, an analog/ASCII interface, a pressure transducer, and fluid conduits.
  • Communication with the implanted pump is achieved via percutaneous needle puncture of the drug reservoir refill septum.

Related Experiment Videos

  • Pump performance is assessed by incrementally filling the reservoir and simultaneously measuring reservoir pressure.
  • Main Results:

    • The apparatus generates a real-time plot of pressure versus volume.
    • This pressure-volume plot quickly identifies or rules out reservoir pressure as a failure mode.
    • Diagnostic runs are saved for archival purposes.

    Conclusions:

    • The portable pump interrogation apparatus offers a practical solution for diagnosing implantable drug infusion pump failures.
    • This device is expected to be as beneficial as the previously developed benchtop apparatus for pump technology development and clinical use.