Portable data acquisition and control apparatus for implanted drug infusion pump interrogation
A S Cress1, B D Wigness, F D Dorman
1Division of Biomedical Engineering, University of Minnesota, Minneapolis 55455.
Summary
Diagnosing implantable drug infusion pump failures is crucial. A new portable device monitors pump performance by measuring reservoir pressure during refills, simplifying failure detection.
Area of Science:
- Biomedical Engineering
- Medical Device Technology
- Drug Delivery Systems
Background:
- Implantable drug infusion pumps are widely used clinically.
- Methods for diagnosing suspected pump failures are needed.
- Previous benchtop apparatus was cumbersome for in vivo monitoring.
Purpose of the Study:
- To develop a portable apparatus for diagnosing implantable drug infusion pump failures.
- To provide a simpler and more accessible method for in vivo pump monitoring.
Main Methods:
- A portable interrogation unit was created, comprising a laptop computer, data acquisition software, an analog/ASCII interface, a pressure transducer, and fluid conduits.
- Communication with the implanted pump is achieved via percutaneous needle puncture of the drug reservoir refill septum.
- Pump performance is assessed by incrementally filling the reservoir and simultaneously measuring reservoir pressure.
Main Results:
- The apparatus generates a real-time plot of pressure versus volume.
- This pressure-volume plot quickly identifies or rules out reservoir pressure as a failure mode.
- Diagnostic runs are saved for archival purposes.
Conclusions:
- The portable pump interrogation apparatus offers a practical solution for diagnosing implantable drug infusion pump failures.
- This device is expected to be as beneficial as the previously developed benchtop apparatus for pump technology development and clinical use.