Related Experiment Video
Updated: Oct 20, 2026

Characterizing Bacterial Volatiles using Secondary Electrospray Ionization Mass Spectrometry (SESI-MS)
Published on: June 8, 2011
Evaluation of Emit II reagents on the Chem 1
R C Sreenivasam1, T C Sneath, N C Jain
1National Toxicology Laboratories, Inc., Bakersfield, CA 93304.
Abstract:
An evaluation study of Syva Emit II reagents using the Chem 1 was performed for the following drugs: barbiturates, benzodiazepines, cannabinoids, benzoylecgonine, opiates, and phencyclidine. The Emit II reagents (100-mL bottles) were reconstituted to 70 mL and evaluated against the Emit d.a.u. reagents. A minimum of 446 samples were run for each drug. For all drugs tested, there were a total of 11 discordant results between the two reagents. The Emit II reagent was found to be correct on 8 of the 11 discordances after retesting by FPIA or GC/MS. The CV of within-run and day-to-day precision of the Emit II was 1.8% or less and 12.3% or less, respectively.
More Related Videos
10:13Eye Irritation Test (EIT) for Hazard Identification of Eye Irritating Chemicals using Reconstructed Human Cornea-like Epithelial (RhCE) Tissue Model
Published on: August 23, 2015
11:44Using Cyclic Voltammetry, UV-Vis-NIR, and EPR Spectroelectrochemistry to Analyze Organic Compounds
Published on: October 18, 2018
Related Concept Videos
Chemical Ionization (CI) Mass Spectrometry
Gas Chromatography: Types of Detectors-II