Dynamic range requirements in digital mammography

A D Maidment1, R Fahrig, M J Yaffe

  • 1Department of Medical Biophysics, University of Toronto, Ontario, Canada.

Medical Physics
|November 1, 1993
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Comparison of radiation dose between 2D digital stereotactic versus digital breast tomosynthesis-guided breast biopsies.

European journal of radiology·2020
Same author

Breast cancer screening policy-good science should trump bad politics.

Current oncology (Toronto, Ont.)·2020
Same author

Three-dimensional tumor visualization of invasive breast carcinomas using whole-mount serial section histopathology: implications for tumor size assessment.

Breast cancer research and treatment·2019
Same author

Investigation of combined kV/MV CBCT imaging with a high-DQE MV detector.

Medical physics·2018
Same author

Factors influencing survival among patients with HER2-positive metastatic breast cancer treated with trastuzumab.

Breast cancer research and treatment·2018
Same author

Motion compensation for cone-beam CT using Fourier consistency conditions.

Physics in medicine and biology·2017
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

Digital mammography requires approximately 3,100 gray levels for optimal imaging, with tungsten (W) spectra being preferable over molybdenum (Mo) due to detector dynamic range needs. Scattered radiation significantly reduces discernable gray levels.

Area of Science:

  • Medical Physics
  • Radiology
  • Imaging Science

Background:

  • Digital mammography requires a high dynamic range and number of gray levels for accurate breast tissue visualization.
  • Understanding the interplay between X-ray spectra, detector characteristics, and image quality is crucial for optimizing mammography systems.

Purpose of the Study:

  • To evaluate the dynamic range and gray levels necessary for digital mammography using an energy transport model.
  • To compare the performance of molybdenum (Mo) and tungsten (W) target spectra under various imaging conditions.

Main Methods:

  • An energy transport model was employed, incorporating Mo and W target spectra.
  • The model accounted for energy-dependent attenuation in elemental filters, breast tissue, and phosphor screens.

Related Experiment Videos

  • The impact of detector noise and coupling efficiency on image quality was assessed.
  • Main Results:

    • Approximately 3,100 gray levels (12 bits/pixel) are needed for ideal detectors with a 40 kVp W spectrum.
    • Tungsten (W) spectra are preferred over molybdenum (Mo) for digital mammography due to lower detector dynamic range requirements.
    • Scattered radiation significantly reduces discernable gray levels, by 23% for a 5 cm breast and 34% for an 8 cm breast without scatter rejection.

    Conclusions:

    • Tungsten (W) target spectra offer advantages for digital mammography regarding detector dynamic range.
    • Detector noise and coupling efficiency impact the number of discernable gray levels, but W spectra remain preferable.
    • Effective scatter rejection is essential to maintain image quality and the number of discernable gray levels in digital mammography.