Possible errors in the measurement of retinal lesions
J V Arnold1, J W Gates, K M Taylor
1Cardiac Surgery Unit, Royal Postgraduate Medical School, London, United Kingdom.
Purpose:
To determine the accuracy of the measurement of structure in the retina from fundus photographs.
Method:
A model eye was developed to measure the retinal image size of the Zeiss fundus camera. A x 10 microscope objective with a focal length of 16.41 mm was used together with a microscope graticule.
Results:
Under emmetropic conditions the difference between the calculated and the measured magnification was 8.6%. The total change in magnification from myopia to hyperopia was -24.63% to +18.1%. When the camera position was altered with respect to the model eye by +/- 5 mm under myopic conditions the change in magnification was -4.5% and +6.8%. In the hyperopic condition the change was -5.6% to +4.6%.
Conclusions:
Measuring the size of structure in the retina from photographs may therefore prove to be inaccurate but in some circumstances determining the percentage change in size may prove more reliable.