The esophageal detector device. Does it work?
Related Experiment Video
Updated: Nov 8, 2026

Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
L Zaleski1, D Abello, M I Gold
1Department of Anesthesiology, University of Miami School of Medicine, Florida.
Background:
The esophageal detector device (EDD) is a diagnostic tool for confirmation of tracheal intubation. Capnography is the accepted standard for such confirmation. The purpose of this investigation was to determine whether results using the EDD and capnography agree.
Methods:
Five hundred patients were divided into three separate studies. In study 1, with 300 consecutive patients, tracheal intubation was performed and tested with the EDD followed by capnography. In study 2, 100 patients had the esophagus intentionally intubated, and confirmation was tested similarly. The tube was then removed and the trachea intubated, and testing followed. Study 3 involved 100 patients and used a double-blind, randomized design. The tube was intentionally inserted into either the esophagus (n = 5) or trachea (n = 49), and testing followed.
Results:
In study 1, the compressed EDD bulb reinflated 270 times and always agreed with capnography; in 20 of the 270 subjects (7%) bulb reinflation was delayed, taking from 5-30 s. In 30 instances, the bulb remained compressed, and there was no capnogram indicating esophageal intubation. In study 2, regardless of esophageal or tracheal intubation, agreement between EDD and capnogram was 100%. In study 3, the agreement between the two detecting instruments was 100%, but reinflation of the EDD bulb was delayed in 4% of tracheal intubations. In the 500 patients studied, results from the EDD and capnogram always agreed, but in 6% of all tracheal intubations, the EDD bulb inflated slowly. Of 181 esophageal intubations, the results from the EDD and capnogram always agreed, i.e., there was no reinflation or capnogram. The sensitivity, specificity, and predictive value for the EDD in all of these studies was 100%.
Conclusions:
The EDD is a valuable diagnostic technique for confirming tracheal intubation. Results using EDD agree with results using capnography; in 6% of instances there is a slow reinflation; and where there is no capnography, such as on hospital wards, EDD may be a useful diagnostic tool.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Esophagus
The movement of edibles from the pharynx into the esophagus is facilitated by the upper esophageal sphincter, which is formed primarily by the...
Barrett Esophagus-II: Clinical Manifestations and Management
To diagnose Barrett's esophagus, healthcare providers often recommend an endoscopy for those showing symptoms of acid reflux. The procedure entails...
Barrett Esophagus-I: Introduction
This constant acid exposure transforms the esophagus's pink mucosal lining (stratified squamous epithelium) into a type of lining more similar...