The effect of sensor malpositioning on pulse oximeter accuracy during hypoxemia
S J Barker1, J Hyatt, N K Shah
1University of California, Irvine Medical Center, Department of Anesthesiology, Orange 92613-1491.
Anesthesiology
|August 1, 1993
Summary
Improperly positioned pulse oximeter sensors can lead to inaccurate oxygen saturation readings, especially during hypoxemia. Frequent sensor checks are crucial to ensure reliable monitoring of arterial oxygen saturation (SaO2).
Area of Science:
- Medical Devices
- Clinical Monitoring
- Respiratory Physiology
Background:
- Pulse oximeters are widely used for non-invasive monitoring of arterial oxygen saturation (SpO2).
- Previous research indicated potential inaccuracies with malpositioned sensors in normoxemic individuals.
- The performance of pulse oximeters with malpositioned sensors during hypoxemia remained uninvestigated.
Purpose of the Study:
- To evaluate the behavior of various pulse oximeters with both correctly and incorrectly positioned sensors.
- To assess oximeter performance across a broad spectrum of arterial oxygen saturation (SaO2) levels, including hypoxemia.
- To identify potential discrepancies between measured SpO2 and actual SaO2 under different sensor placement conditions.
Main Methods:
- Eight different pulse oximeters were tested on 12 healthy volunteers.
- Five oximeters were intentionally used with malpositioned sensors.
- Subjects underwent controlled breathing of oxygen-nitrogen mixtures to achieve SaO2 levels ranging from 100% down to 70%, with arterial blood gas analysis for calibration.
Main Results:
- Malpositioned sensors significantly altered pulse oximeter performance, with variations observed between manufacturers and models.
- Some devices consistently underestimated SpO2, while others showed a trend of underestimation at high SaO2 and overestimation at low SaO2.
- Analysis revealed a reduced sensitivity of SpO2 to SaO2 changes (decreased slope) in many cases, alongside significant inter-subject variability.
Conclusions:
- Sensor malpositioning substantially alters pulse oximeter calibration curves.
- In hypoxemic conditions, inaccurate readings due to malpositioned sensors can mask severe oxygen desaturation, leading to underestimation of hypoxemia.
- Recommendations include frequent verification of sensor placement and avoidance of difficult-to-access sensor sites.