Comparative evaluation of three flow cytometers for reticulocyte enumeration
M Van Petegem1, R Cartuyvels, P de Schouwer
1University Hospital Leuven, Department of Haematology, Belgium.
Clinical and Laboratory Haematology
|January 1, 1993
Summary
Flow cytometry is replacing manual methods for counting reticulocytes. Three instruments (FACScan, Profile II, Cytoron Absolute) showed good comparability for reticulocyte enumeration, suitable for routine use.
Area of Science:
- Hematology
- Clinical Pathology
- Flow Cytometry
Background:
- Manual microscopic methods for reticulocyte enumeration are being superseded by flow cytometry.
- Flow cytometry offers automated and more precise analysis of red blood cell populations.
Purpose of the Study:
- To evaluate and compare the performance of three different flow cytometers for reticulocyte enumeration.
- To assess the comparability of reticulocyte counts obtained from Becton Dickinson FACScan, Coulter EPICS Profile II, and Ortho Cytoron Absolute instruments.
Main Methods:
- Evaluation of three flow cytometers (FACScan, EPICS Profile II, Cytoron Absolute) using thiazole orange dye.
- Analysis of 30,000 cells per sample, with manual gating of the erythroid population for fluorescence evaluation.
- Orthogonal regression analysis and paired t-tests were used to compare instrument performance on 100 healthy donor and 108 patient samples.
Main Results:
- All three instruments demonstrated good linearity and precision in reticulocyte counting.
- No significant carry-over was observed between samples.
- While statistically significant differences were noted between the Cytoron Absolute and the other two instruments, overall mutual comparability was good.
Conclusions:
- The Becton Dickinson FACScan, Coulter EPICS Profile II, and Ortho Cytoron Absolute are comparable for routine reticulocyte enumeration.
- Despite minor statistical variations, these flow cytometers provide reliable results for clinical practice.
- Flow cytometry offers an efficient and accurate alternative to manual reticulocyte counting.