Film density calibration for computed radiography systems: is the standard three-point procedure accurate?
1Department of Radiology, College of Medicine, J. Hillis Miller Health Center, University of Florida, Gainesville 32610-0374.
Radiology
|September 1, 1993
Abstract:
With computed radiography (CR) systems, laser printer-film processor combinations are normally checked daily by using test film three-point density measurements. In this study, the authors show that the recommended three-point quality assurance procedure can show satisfactory results even though the CR film output behaves in an anomalous manner. This problem can be corrected by using a 16-point calibration procedure. Regular 16-point calibrations should be performed to ensure satisfactory CR system performance.