Phantom pacemaker programming

A Fieldman, R J Dobrow

    Pacing and Clinical Electrophysiology : PACE
    |April 1, 1978
    PubMed
    Summary
    This summary is machine-generated.

    Pacemaker programming errors, including dysprogramming and misprogramming, can occur but were not clinically detrimental in this study. Awareness of these potential hazards during pacemaker follow-up is crucial for patient safety.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Low pressure radiofrequency balloon angioplasty: evaluation in porcine peripheral arteries.

    Journal of the American College of Cardiology·1993
    Same author

    Effects of different dosages and modes of sodium bicarbonate administration during cardiopulmonary resuscitation.

    The American journal of emergency medicine·1992
    Same author

    Diltiazem improves resuscitation from experimental ventricular fibrillation in dogs.

    Critical care medicine·1992
    Same author

    Evaluation and comparison of the adverse effects of streptokinase and alteplase.

    Pharmacotherapy·1992
    Same author

    The effect of CPR on plasma diltiazem concentrations in dogs.

    Annals of emergency medicine·1991
    Same author

    Epinephrine versus methoxamine in survival postventricular fibrillation and cardiopulmonary resuscitation in dogs.

    Critical care medicine·1989
    Same journal

    A Case of Permanent Pacemaker Implantation via the Epicardial Approach Using the 3830 Lead in an 11-Day-Old Neonate (With Follow-Up of the Above Case).

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Cryoballoon Versus Radiofrequency Ablation for Persistent Atrial Fibrillation: Meta-Analysis of Randomized Trials.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Tilt Test Duration in Suspected Vasovagal Syncope: Temporal Patterns and Diagnostic Yield in Patients From Central China.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Combined Leadless Pacing and Subcutaneous ICD Therapy in Long QT Syndromes.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Association of Anesthesia Modality With Procedural Parameters and Clinical Outcomes in PVI for Atrial Fibrillation.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Zero-Fluoroscopy Cryoballoon Ablation for Paroxysmal Atrial Fibrillation in a Patient With Dextrocardia: A Case Report.

    Pacing and clinical electrophysiology : PACE·2026
    See all related articles

    Area of Science:

    • Biomedical Engineering
    • Cardiology
    • Medical Device Technology

    Background:

    • Programmable pacemakers are widely used for cardiac rhythm management.
    • Spurious programming, including dysprogramming and misprogramming, poses potential risks.
    • Clinical reports of such events have been rare despite device prevalence.

    Purpose of the Study:

    • To investigate the incidence and mechanisms of spurious programming in pacemakers.
    • To identify potential causes of dysprogramming and misprogramming.
    • To alert healthcare professionals to potential pacemaker programming anomalies.

    Main Methods:

    • Routine follow-up of 95 implanted Cordis Omni-Stanicor generators.
    • Observation of dysprogramming events during magnet application.

    Related Experiment Videos

  • In vitro studies to assess magnetically induced reed switch vibration.
  • Analysis of misprogramming instances related to programmer faults.
  • Main Results:

    • Three instances of dysprogramming were observed in vivo.
    • Magnetically induced reed switch vibration was found in 9 of 6,680 units in vitro.
    • Numerous misprogramming events were attributed to faulty programmer emission counts.
    • No clinically detrimental anomalies were identified in this series.

    Conclusions:

    • Magnet application can induce pacemaker dysprogramming due to reed switch vibration.
    • Faulty programmer emission counts are a common cause of misprogramming.
    • While not clinically detrimental in this study, these phenomena warrant awareness and caution in pacemaker management.