Performance assessment and quality control in MRI by Eurospin test objects and protocols

R A Lerski1, J D de Certaines

  • 1Department of Medical Physics, Ninewells Hospital and Medical School, Dundee, UK.

Magnetic Resonance Imaging
|January 1, 1993
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Magnetic resonance imaging texture analysis classification of primary breast cancer.

European radiology·2015
Same author

Quantitative analysis of cardiac left ventricular variables obtained by MRI at 3 T: a pre- and post-contrast comparison.

The British journal of radiology·2012
Same author

Is biomedical nuclear magnetic resonance limited by a revisitable paradigm in physics?

Cellular and molecular biology (Noisy-le-Grand, France)·2005
Same author

1H NMR structural characterization of the cytochrome c modifications in a micellar environment.

Biochemistry·2003
Same author

MRI texture analysis on texture test objects, normal brain and intracranial tumors.

Magnetic resonance imaging·2003
Same author

Is magnetic resonance imaging texture analysis a useful tool for cell therapy in vivo monitoring?

Anticancer research·2002
Same journal

Repeatability of an MRI protocol for generating habitats based on cellularity, perfusion, and hypoxia in a murine model of glioma.

Magnetic resonance imaging·2026
Same journal

Association between CMR-derived pulmonary artery pulse wave velocity and pulmonary risk factors: the multi-ethnic study of atherosclerosis COPD study.

Magnetic resonance imaging·2026
Same journal

Systematic comparison of MPRAGE and BRAVO T1-weighted MRI pulse sequences and brain morphometry in high-risk young adults.

Magnetic resonance imaging·2026
Same journal

Foot dynamic contrast-enhanced MRI for assessing microcirculatory changes after endovascular therapy in peripheral artery disease: A prospective pilot study.

Magnetic resonance imaging·2026
Same journal

Reconstruction of MRI from undersampled k-spaces of double-contrast volume acquisitions using deep neural networks.

Magnetic resonance imaging·2026
Same journal

Radiofrequency-induced heating safety of brain MRI scans at 7 T in the presence of a shoulder implant.

Magnetic resonance imaging·2026
See all related articles

The Eurospin set provides standardized magnetic resonance imaging (MRI) quality control through test objects and measurements. This ensures consistent MRI performance from acceptance testing to routine quality assurance.

Area of Science:

  • Medical Imaging Physics
  • Biomedical Engineering
  • Quality Control in Healthcare

Background:

  • Magnetic Resonance Imaging (MRI) requires rigorous quality control for accurate tissue characterization.
  • The European Concerted Action "Tissue Characterisation by MRS and MRI" (COMAC-BME) identified the need for standardized MRI quality assessment tools.

Purpose of the Study:

  • To introduce the Eurospin set, a collection of test objects designed for MRI quality control.
  • To detail the history, development, and application of the Eurospin set in MRI systems across Europe.
  • To outline principles for advanced (Stage II) MRI testing and methods for performance quantification.

Main Methods:

  • Design and rigorous testing of the Eurospin set of MRI test objects.
  • Application of test objects across multiple MRI systems throughout Europe.

Related Experiment Videos

  • Introduction of Stage II tests involving invasive system interaction and discussion of simplified performance quantification methods.
  • Main Results:

    • The Eurospin set enables a comprehensive range of measurements for assessing MRI performance.
    • Data from extensive testing across European systems demonstrate the utility and reliability of the Eurospin set.
    • Established protocols for MRI quality control, from initial acceptance testing to ongoing quality assurance.

    Conclusions:

    • The Eurospin set is a validated tool for effective MRI quality control.
    • Standardized testing procedures ensure consistent and reliable MRI performance across different systems.
    • The framework supports a multi-level approach to MRI quality assurance, enhancing diagnostic accuracy.