Blood vessel diameter measurement by ultrasound
Related Experiment Video
Updated: Nov 7, 2026

Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
S Li1, W N McDicken, P R Hoskins
1Department of Medical Physics and Medical Engineering, Royal Infirmary of Edinburgh, UK.
Abstract:
Blood vessel diameter measurement is crucial to volume flow measurement in medical ultrasound. The accuracy of the measurement is examined in this study. Experiments were made on three kinds of tube using four transducers on three commercial duplex ultrasound scanners. The accuracy of the inner diameter measurement is affected by ultrasonic pulse length. By correcting this effect, it has been shown that the accuracy of the measurement can be increased significantly over a range of conditions.