Related Experiment Videos
The relationship between two power-grip testing devices and their utility in physical capacity evaluations
L D Robertson1, C M Mullinax, G R Brodowicz
1Work Fitness Center, Exeter Hospital, New Hampshire 03833.
Summary
This study found that two isometric grip testing devices reliably measure hand strength. Combining data from these devices can help assess functional status and effort consistency in patients with upper-extremity impairments.
Area of Science:
- Rehabilitation Medicine
- Orthopedics
- Physical Therapy
Background:
- Accurate assessment of physical capacity is crucial for diagnosing and managing upper-extremity impairments.
- Evaluating consistency of effort is challenging but vital for reliable physical capacity evaluations (PCEs).
- Existing methods for grip strength assessment may have limitations in capturing functional status and effort consistency.
Purpose of the Study:
- To compare the reliability of two isometric grip testing devices: the Jamar dynamometer and the Isometric Strength Testing Unit (ISTU).
- To determine the efficacy of combining data from these devices for assessing functional status and consistency of effort in PCEs.
- To develop and evaluate a novel index (2-ratio) for functional status and effort consistency.
Main Methods:
- Recruited 26 healthy subjects and 40 patients with upper-extremity impairments.
- Measured maximal isometric grip force using both the Jamar dynamometer and the ISTU.
- Analyzed the relationship between force outputs from the two devices.
- Developed and applied a 2-ratio index using data from both devices.
Main Results:
- High correlation was found between power-grip force outputs from the Jamar dynamometer and the ISTU.
- This indicates the utility of both devices for PCEs involving wrist-hand function.
- A simple 2-ratio index was developed, demonstrating potential for assessing functional status and effort consistency.
Conclusions:
- Isometric grip testing devices demonstrate reliable and related force outputs for wrist-hand involvement.
- Combining data from multiple devices enhances the assessment of functional status and effort consistency.
- The developed 2-ratio index shows promise for improving the sensitivity of PCEs.