Field characteristics of geometric unsharpness due to the x-ray tube focal spot
Related Experiment Video
Updated: Aug 12, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Abstract:
The basic theory of field characteristics is discussed. It is shown that the point spread function and optical transfer function (OTF) of geometric unsharpness at any arbitrary field position can be derived from those at the central beam position. At all locations in the radiation field, the OTF in a direction parallel to the x-ray tube cathode-target axis does not depend on the field position X, which is perpendicular to the tube axis, whereas the OTF in a perpendicular direction does not depend on the field position Y, which is parallel to the tube axis. The theoretical result is used to demonstrate the field characteristic of the uniform, square focal spot.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...