Spectral considerations for absorption-edge fluoroscopy

F Kelcz, C A Mistretta, S J Riederer

    Medical Physics
    |January 1, 1977
    PubMed
    Summary
    This summary is machine-generated.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Quantification of Blood Velocity with 4D Digital Subtraction Angiography Using the Shifted Least-Squares Method.

    AJNR. American journal of neuroradiology·2018
    Same author

    Experimental realization of fluence field modulated CT using digital beam attenuation.

    Physics in medicine and biology·2014
    Same author

    Semi-automated vectorial analysis of anorectal motion by magnetic resonance defecography in healthy subjects and fecal incontinence.

    Neurogastroenterology and motility·2012
    Same author

    Gastric motor disturbances in patients with idiopathic rapid gastric emptying.

    Neurogastroenterology and motility·2011
    Same author

    Application of magnetic resonance imaging to measure fasting and postprandial volumes in humans.

    Neurogastroenterology and motility·2008
    Same author

    3D time-resolved contrast-enhanced cerebrovascular MR angiography with subsecond frame update times using radial k-space trajectories and highly constrained projection reconstruction.

    AJNR. American journal of neuroradiology·2007
    Same journal

    Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

    Medical physics·2026
    Same journal

    Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

    Medical physics·2026
    Same journal

    Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

    Medical physics·2026
    Same journal

    Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

    Medical physics·2026
    Same journal

    Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

    Medical physics·2026
    Same journal

    Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

    Medical physics·2026
    See all related articles

    This study enhances absorption-edge fluoroscopy by optimizing spectral parameters. A 3-kVp, 3-filter technique with logarithmic processing allows imaging iodine and xenon despite significant tissue and bone variations.

    Area of Science:

    • Medical Imaging
    • Radiology
    • Physics

    Background:

    • Absorption-edge fluoroscopy is a valuable imaging technique.
    • Previous methods faced limitations in spectral parameter selection for accurate imaging.
    • Tissue and bone variations complicate contrast agent visualization.

    Purpose of the Study:

    • To provide an overview of spectral parameter selection in absorption-edge fluoroscopy.
    • To present an improved technique for imaging specific elements in the presence of tissue variations.
    • To address limitations of existing methods for bone thickness variations.

    Main Methods:

    • Utilized a 1-kilovolt peak (kVp), 2-filter technique for initial imaging.
    • Incorporated logarithmic signal processing to extend imaging range.

    Related Experiment Videos

  • Developed and simulated a 3-kVp, 3-filter technique with logarithmic signal processing.
  • Main Results:

    • The 1-kVp, 2-filter technique allowed imaging elements like iodine and xenon with +/-2 cm tissue variations.
    • Logarithmic processing improved tissue variation tolerance but not bone variations.
    • The 3-kVp, 3-filter technique with logarithmic processing successfully imaged 1 mg/cm² iodine with >10 cm tissue and >2000 mg/cm² bone variations.

    Conclusions:

    • Optimized spectral parameter selection is crucial for absorption-edge fluoroscopy.
    • The 3-kVp, 3-filter technique with logarithmic signal processing significantly enhances robustness against tissue and bone variations.
    • This advanced technique improves the potential for accurate elemental imaging in complex biological environments.