Effects of x-ray tube-voltage ripple and current mode on patient dose
H Kubota1, Y Ozaki, M Matsumoto
1Information Processing Center, Shiga University, Japan.
Abstract:
This paper proposes a radiographic sensitivity attenuation curve for screen-films systems. This curve expresses the relative inverse value of mAs value needed to give the same photographic density. This curve decreases more slowly with increasing object thickness than the exposure attenuation curve generally used hitherto. By using the radiographic sensitivity attenuation curve and the exposure attenuation curve, the voltage-ripple dependence of mAs values and patient dose for the same photographic density was obtained. As predicted theoretically in a previous paper [Med. Phys. 18, 921 (1991)], it was verified quantitatively the reverse order that larger ripple values resulted in lower patient dose and lower mAs value for the unsaturating current mode at large ripple. This reverse order clearly appeared in two-pulse units. These phenomena were experimentally verified. As a result it was found that in radiographing with a two-pulse unit, larger mA value results in lower patient dose.