MR angiography of the portal and hepatic venous systems: preliminary experience with echoplanar imaging
M A Goldberg1, E K Yucel, S Saini
1Department of Radiology, Massachusetts General Hospital, Boston 02114.
Objective:
The purpose of this study was to evaluate the ability of echoplanar MR angiography to depict the major hepatic and portal venous structures.
Subjects And Methods:
Echoplanar and conventional MR angiographic examinations were performed in 10 subjects (seven healthy volunteers, three patients with focal hepatic lesions). A gradient-recalled echo (GRE) time-of-flight technique (125/10 [TR/TE], 90 degrees flip angle) was used for echoplanar angiography. Eight complete single-excitation images were acquired at each level in 1.5 sec and then collapsed into a single maximal intensity projection. Conventional time-of-flight MR angiography (34/13, 30 degrees flip angle) also was performed. The vascular anatomy from the right atrium to the splenic vein was imaged (6-mm contiguous levels) in three 10.5-sec breath-holds with echoplanar imaging, as compared with seven 11.5-sec breath-holds with conventional MR angiography. Echoplanar and conventional images were compared quantitatively and qualitatively.
Results:
Echoplanar imaging was 61% faster than conventional MR angiography. Vessel-to-liver signal-intensity ratios were significantly higher for echoplanar imaging (p < .0001), signal-to-noise ratios were significantly higher for conventional MR angiography (p < .0001), and contrast-to-noise ratios were comparable. Qualitatively, echoplanar imaging and conventional MR angiography provided similar anatomic information about the hepatic and portal veins.
Conclusion:
Angiograms of the hepatic and portal venous systems that are of diagnostic quality can be acquired much more quickly with echoplanar imaging than with conventional MR angiography.