Related Experiment Video
Updated: Aug 11, 2026

A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Measurement of activity in anthropomorphic organ phantoms
T E Hui1, L L Nichols, P C Olsen
1Health Physics Department, Pacific Northwest Laboratory, Richland, WA 99352.
Abstract:
This paper describes two nondestructive measurement techniques for determining the radioactivity in a homogeneous organ phantom with an acceptable error and the capability of being traceable to the National Institute of Standards and Technology. These two techniques are based on a method developed by Robley D. Evans in 1937 for measuring the amount of radium deposited in a living person. There are two significant improvements in the new techniques: 1) the radially-dependent error is eliminated, and 2) the effect of self absorption in the unknown body is measured and taken into account. The first assay method is a single-source technique involving four measurements and requiring only one standard source, the second one is a double-source technique involving six measurements and requiring two unequal standard sources. Three pairs of lung phantoms radiolabeled with 241Am, 137Cs, and 154Eu were measured with both the single- and double-source techniques. The results imply that the self-attenuation of photons within the organ phantom cannot be neglected, especially at low energies. Finally, potential applications (other than the calibration of bioassay phantoms of the single- and double-source techniques) and their limitations are discussed.