Maintenance of defibrillators in a state of readiness

R D White1

  • 1Department of Anesthesiology, Mayo Clinic, Rochester, Minnesota.

Insights

Adverse incidents with external defibrillators are often due to improper operation and maintenance. The US Food and Drug Administration (FDA) developed checklists to improve defibrillator safety and battery care.

Area of Science:

  • Medical device safety
  • Public health surveillance

Background:

  • The US Food and Drug Administration (FDA) established the Medical Device Reporting system in 1984 to track adverse events associated with medical devices.
  • External defibrillators, critical in emergency cardiac care, have been frequently implicated in adverse incident reports.

Observation:

  • An FDA study investigated defibrillator user training, device maintenance, and battery upkeep across five states.
  • Inspections and testing of defibrillators and batteries were conducted to evaluate their maintenance status.

Findings:

  • A significant majority of adverse incidents were linked to improper defibrillator operation and maintenance practices.
  • Battery failure was identified as a recurring and preventable issue, necessitating a focus on proper battery maintenance (nickel-cadmium and sealed lead-acid).

Implications:

  • The FDA Center for Devices and Radiological Health initiated an educational program targeting defibrillator operators and maintenance personnel.
  • Checklists for manual and automated defibrillators were created to address common device and user-related problems, promoting regular use in all settings.