A simple method of producing depth ionization data for electron energy constancy check
M K Islam1, H Rashid, H Gaballa
1Department of Clinical Physics, Princess Margaret Hospital, Toronto, Canada.
Abstract:
A simple method has been developed to reproduce depth ionization data of electron beams for energy determination. The method utilizes a simple set of equipment, a combination of a specially designed wedge-shaped polystyrene phantom and a linear array of detectors, to collect the necessary data. The wedge-shaped phantom provides varying depths to various detectors in the array. The ionization readings received from the detectors were corrected for off-axis ratio and plotted against corresponding ray-line depths to produce depth ionization curves. The instrument setup was fast and simple. The relevant data, for a high-energy linear accelerator with multiple electron energies, were collected in minutes. The depths of 80% and 50% ionization determined by this method were found to differ by 2 and 3 mm, respectively, at the most, with those determined by a conventional method.