Automated blood pressure measurements in laboring women: are they reliable?
G F Marx1, S S Schwalbe, E Cho
1Department of Anesthesiology, Albert Einstein College of Medicine, Bronx, NY 10461.
Objective:
In laboring women a consistent difference has become evident between measurements obtained with an automated blood pressure device and those obtained with the auscultatory method. A prospective study was designed to assess the concordance of these two methods.
Study Design:
Three sets of brachial blood pressure measurements were made by both oscillatory and auscultatory techniques in 30 women in labor, 20 term pregnant women not in labor, and 20 nonpregnant volunteers.
Results:
In the nonlaboring women and the nonpregnant controls there was satisfactory agreement between the results of the two methods of measurement. In the parturients systolic pressures were consistently and significantly higher and diastolic pressures consistently and significantly lower with the oscillatory compared with the auscultatory method, but mean arterial pressures were not different.
Conclusion:
In laboring women there is a discrepancy between systolic and diastolic pressures obtained by the auscultatory versus the oscillatory method of measurement, although mean pressures are not significantly different. We suggest that during labor the diagnoses of hypertension and hypotension be based on the mean rather than the systolic or diastolic pressure.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure