Comparison of portable emergency ventilators using a lung model

L Attebo1, M Bengtsson, A Johnson

  • 1Department of Anaesthesiology, University Hospital, Linköping, Sweden.

British Journal of Anaesthesia
|March 1, 1993
PubMed
Summary

Five emergency ventilators were tested using a lung model. Four ventilators (MEDUMAT Elektronik, MEDUMAT Variabel, OSIRIS, and OXYLOG) demonstrated acceptable performance in simulated obstructive lung disease.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Defective B cell development in Snell dwarf (dw/dw) mice can be corrected by thyroxine treatment.

Journal of immunology (Baltimore, Md. : 1950)·1996
Same author

The carboxyl terminus of GLUT4 contains a serine-leucine-leucine sequence that functions as a potent internalization motif in Chinese hamster ovary cells.

The Journal of biological chemistry·1996
Same author

Safe feeding practices for infants and young children.

Journal of paediatrics and child health·1996
Same author

Clinical associations and time of onset of cerebral white matter damage in very preterm babies.

Archives of disease in childhood. Fetal and neonatal edition·1996
Same author

Frequency-specific auditory brainstem responses in adults with sensorineural hearing loss.

Audiology : official organ of the International Society of Audiology·1996
Same author

Gun acquisition and use by juvenile offenders.

JAMA·1996

Area of Science:

  • Biomedical Engineering
  • Respiratory Medicine
  • Medical Device Technology

Background:

  • Emergency ventilators are critical for respiratory support in critical care.
  • Evaluating ventilator performance under simulated disease conditions is essential for clinical safety.
  • Obstructive lung disease presents unique challenges for mechanical ventilation.

Purpose of the Study:

  • To assess the performance of five emergency ventilators.
  • To compare preset and measured minute volumes.
  • To evaluate gas distribution in a simulated obstructive lung disease model.

Main Methods:

  • A lung model with adjustable compliance and resistance was utilized.
  • Five ventilators (MEDUMAT Elektronik, MEDUMAT Variabel, OSIRIS, OXYLOG, rescuPAC 2DM) were tested.

Related Experiment Videos

  • Minute volume accuracy and gas distribution were measured and compared.
  • Main Results:

    • MEDUMAT Elektronik, MEDUMAT Variabel, OSIRIS, and OXYLOG ventilators showed acceptable performance.
    • The rescuPAC 2DM ventilator's performance was not explicitly detailed as acceptable in the abstract.
    • Ventilator performance was evaluated against preset parameters and simulated patient conditions.

    Conclusions:

    • MEDUMAT Elektronik, MEDUMAT Variabel, OSIRIS, and OXYLOG are suitable for use in simulated obstructive lung disease.
    • The lung model effectively simulated key aspects of obstructive lung disease for ventilator testing.
    • Further evaluation may be warranted for ventilators not meeting performance criteria.