Related Experiment Videos
Tomogram degradation due to eye shielding and collimation
Radiology
|April 1, 1977
Abstract:
Phantom-image, line-pair resolution and anterior eye exposure have been measured to determine the optimum collimation and eye shielding for petrous temporal tomography. Small round collimation and 0.5-1.0-mm lead equivalent eye shields are recommended. This reduces eye exposure to 10% of the usual dose without changing line-pair resolution.