"Pseudo auto-PEEP"? A new cause for discrepancy between the end-expiratory occlusion plateau pressure and airway opening pressure
Related Experiment Video
Updated: Jul 16, 2026

Image Acquisition using Portable Sonography for Emergency Airway Management
Published on: September 28, 2022
Abstract:
The purpose of this study is to describe an unexpected degree of differences between expiratory occlusion plateau pressure (EPO) and airway opening pressure (Pawo) measured level of intrinsic positive end-expiratory pressure above externally applied (auto-PEEP) that was found in six critically ill patients. In six patients (ten studies), the presence and degree of auto-PEEP found during the EPO maneuvers was not confirmed by Pawo measurements. In five studies, flow tracings showed prolonged near zero flow toward end expiration and a slow rise to plateau during the EPO maneuver. Because of the static nature of the EPO determination, a rise in pressure could conceivably be caused by the presence of subcutaneous or mediastinal emphysema and retrograde flow across the airway defect during exhalation. We suggest another cause for auto-PEEP-dynamic hyperinflation from extrapulmonary flow limitation.
Related Concept Videos
Measurement of Blood Pressure
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Atelectasis II: Pathophysiology