Surface dose perturbation due to air gap between patient and bolus for electron beams
1Department of Radiation Oncology, E. W. Sparrow Hospital, Lansing, Michigan 48909.
Abstract:
The effect of air gaps under bolus material on the surface dose for electron beams is investigated. Dose measurements were performed for air gaps from 0.0 to 3.0 cm and bolus thicknesses of 0.5 and 1.0 cm using the various electron energies and cone sizes available on an electron linear accelerator. Our results show that the surface dose decreases for lower electron energies, smaller field sizes, thicker boluses, and larger air gaps.