Calculation of volume flow rate by the proximal isovelocity surface area method: simplified approach using color
T Utsunomiya1, R Doshi, D Patel
1Department of Medicine, University of California, Irvine, Orange.
Objectives:
The goal of this study was to develop an accurate, simplified proximal isovelocity surface area (PISA) method for calculating volume flow rate using lower blue-red interface velocity produced by a color Doppler zero baseline shift technique.
Background:
The Doppler color proximal isovelocity surface area method has been shown to be accurate for calculating the volume flow rate (Q) across a narrowed orifice by the formula Q = PISA x Blue-red interface velocity. A hemispheric model is generally used to calculate proximal isovelocity surface area (PISA = 2 pi a2, where a = the radius corresponding to the blue-red interface velocity). Although a hemispheric model is simple, requiring measurement of one radius, it may underestimate the actual volume flow rate because, in the general case, the shape of a proximal isovelocity surface area is hemielliptic. Although a hemielliptic model is generally more accurate for calculating proximal isovelocity surface area, it is more complex, requiring measurement of two orthogonal radii.
Methods:
Sixteen in vitro constant flow model studies were performed using planar circular orifices (diameter range 6 to 16 mm). The blue-red interface velocity was changed from 3 to 54 cm/s using color Doppler zero baseline shift.
Results:
1) With decreasing blue-red interface velocity, the size of the proximal isovelocity surface area was increased, and its shape changed from hemielliptic to hemispheric. 2) With the blue-red interface velocity in the range 11 to 15 cm/s, the proximal isovelocity surface area became nearly hemispheric; however, it was difficult to determine the blue-red interface radius at a blue-red interface velocity < 10 cm/s because of interface fluctuations. 3) Calculated volume flow rate using the hemispheric proximal isovelocity surface area model with a single radius was relatively accurate at a blue-red interface velocity of 11 to 15 cm/s (mean percent difference from actual volume flow rate was -3.6%).
Conclusions:
Because the shape of the proximal isovelocity surface area is nearly hemispheric at a blue-red interface velocity of 11 to 15 cm/s, volume flow rate can be accurately calculated in this proximal isovelocity surface area interface velocity range (produced by zero baseline shift) by measuring a single-interface radius. This approach should be clinically useful for calculating the volume flow rate across stenotic and regurgitant valves and across shunt defects.