Pathology of silicone intraocular lenses in human eyes obtained postmortem

C A Blotnick1, T P Powers, T Newland

  • 1Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston 29425, USA.

The increased use of phacoemulsification and small incision surgery has led to interest in a soft intraocular lens (IOL) for implantation after cataract extraction. We present the pathologic findings of seven eyes obtained postmortem from five patients with silicone IOLs. Length of implantation ranged from 6 weeks to 13 months. There was symmetric bag-bag fixation in all cases. Two IOLs had distorted and compressed polypropylene loops, resulting in decentration. The silicone IOL appeared to be well tolerated in all cases.