Related Experiment Video
Updated: Aug 4, 2026

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
The Demers atrial catheter: experience with a single-lumen silicone catheter as short- and long-term access for
M Zieschang1, B Erben, D Höffler
1Department of Internal Medicine III, Städtische Kliniken Darmstadt, Germany.
Insights
The Demers catheter, a silicone atrial catheter with a dacron cuff, provides hemodialysis access. Complications led to explantation in 20% of cases, with malfunction and infection being primary reasons.
Area of Science:
- Nephrology
- Vascular Access
- Medical Devices
Background:
- The Demers catheter is a silicone atrial catheter with a dacron cuff.
- It is utilized for both short-term and long-term hemodialysis access.
Purpose of the Study:
- To retrospectively analyze the outcomes of Demers catheter implantations.
- To evaluate the efficacy and complication rates of the Demers catheter for hemodialysis access.
Main Methods:
- Retrospective analysis of 404 Demers catheter implantations in 316 patients.
- Data collected from January 1, 1987, to May 31, 1991.
- Categorization of use into short-term (< 91 days) and long-term (> 90 days).
Main Results:
- Median catheter lifespan was 87 days overall.
- For long-term use, median lifespan was 150 days per catheter and 220 days per patient.
- Catheter malfunction occurred every 876 days; 20% of catheters were explanted due to complications (42 malfunction, 22 infection).
Conclusions:
- The Demers catheter serves as a viable option for hemodialysis access.
- Complication rates, including malfunction and infection, necessitate careful monitoring and management.
- Further research may explore strategies to mitigate these complications for improved patient outcomes.
Abstract:
The Demers catheter is a silicone atrial catheter with a dacron cuff used as short- and long-term access for hemodialysis. It was implanted in 316 patients between January 1, 1987 through May 31, 1991. Data on these implantations were retrospectively analyzed and are reported here. Follow-up and analysis was possible in 404 of 417 Demers catheter implantations. The mean age of patients in this study was 61 years. For short-term use (< 91 days) 153 catheters were implanted in 135 patients, for long-term use (> 90 days) 251 catheters in 181 patients. The median life span of all 404 catheters was 87 days. The median life span for long-term use was 150 days (2-1302) per catheter and 220 days (92-1717) per patient. Catheter malfunction (arterial blood flow < 200 ml) was encountered every 876 days in the running time of the catheter. Twenty percent (n = 80) of the catheters were explanted because of complications. These included 42 cases of catheter malfunction and 22 cases of infection.
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization
Peritoneal Dialysis I: Introduction and Procedure
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications