Related Experiment Video
Updated: Aug 3, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Stabilization of the short sagittal split osteotomy: in vitro testing of different plate and screw configurations
1Clinic for Reconstructive Surgery, University Hospital Basel, Switzerland.
Abstract:
Using a 3-dimensional in vitro model, the stability of different types of osteosynthesis for the short sagittal osteotomy was tested. The following four groups of plate and screw configurations were evaluated: Group I: Fixation with miniplates using monocortical screws only, Group II: Fixation with miniplates, but with two of the screws engaging both fragments, Group III: Same as group II with an additional position screw, Group IV: Fixation with 3 position screws. The stability obtained with miniplate fixation using monocortical screws only (group I) was by a factor of 2.9 less than position screw fixation (group IV), which is considered to be an approved standard. In order to increase the stability of miniplate fixation, the screws in the area of overlapping bone should engage both fragments.