[Results of a study of quality control of diagnostic ultrasound equipment]
1Institut für Biomed. Technik und Physik, Wien.
Aims:
In this paper results of a quality control programme for B-mode ultrasound equipment used in clinical routine are presented. Over a period of 18 months 5 modern scanners with altogether 19 scanheads were periodically controlled.
Methods:
The methods used for measuring the various performance parameters of an ultrasound equipment followed closely standard methods. All measurements were carried out with commercially available tissue-mimicking phantoms, and documented in a uniform protocol.
Results:
Based on the results it was found that 11 of the scanheads did not operate correctly; 2 additional ones were severely defective. Only 6 scanheads were without loss of image quality or function. For the 13 malfunctioning scanheads adjustments, repair or even an exchange was necessary.
Conclusion:
The study demonstrates that quality control performed periodically is necessary to ensure optimal image quality and safe operation of the equipment.