Related Experiment Video
Updated: Aug 9, 2026

05:46
Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile
Published on: September 20, 2024
Intraoperative semiquantitative keratometry using the keratoscopic astigmatic ruler
1Department of Ophthalmology, University of New South Wales, Australia.
Journal of Cataract and Refractive Surgery
|November 1, 1995
Summary
A new keratoscopic astigmatic ruler accurately measures corneal astigmatism during surgery. This handheld device ensures correct measurement distance, improving surgical outcomes for astigmatism correction.
Area of Science:
- Ophthalmology
- Optics
- Medical Devices
Background:
- Intraoperative keratometry is crucial for managing corneal astigmatism via suture tension.
- Hand-held keratometers risk errors due to variable measurement distances affecting image shape.
- Accurate measurement requires maintaining a calibrated distance from the astigmatic cornea.
Purpose of the Study:
- To design and evaluate a novel hand-held keratometer for assessing corneal astigmatism magnitude.
- To improve the accuracy of intraoperative astigmatism measurement.
- To provide surgeons with a reliable tool for surgical decision-making.
Main Methods:
- Development of the keratoscopic astigmatic ruler, a hand-held device with a rack of elliptical keratoscopic lenses.
- The device uses inexpensive, autoclavable, injection-molded plastic.
- Incorporation of a sight mechanism to ensure correct measurement distance and a bright corneal reflection.
Main Results:
- The keratoscopic astigmatic ruler generates a clear aerial image above the cornea.
- The design facilitates easy determination of corneal reflection for astigmatism assessment.
- The integrated sight aids in maintaining the calibrated distance for accurate measurements.
Conclusions:
- The keratoscopic astigmatic ruler is an effective tool for measuring corneal astigmatism intraoperatively.
- Its design addresses the distance-dependent error inherent in traditional hand-held keratometers.
- This device offers a simple, accurate, and cost-effective solution for ophthalmic surgeons.