Comparative study of power-based versus mean frequency-based transcranial color-coded duplex sonography in normal
R W Baumgartner1, C Schmid, I Baumgartner
1Department of Neurology, University of Bern, Inselspital, Switzerland.
Stroke
|January 1, 1996
Summary
Power-based transcranial color-coded duplex sonography (p-TCCD) showed reliable velocity measurements in major cerebral arteries but did not visualize cerebral parenchyma. This new ultrasonic method has limitations compared to frequency-based TCCD (f-TCCD).
Area of Science:
- Neurosonology
- Cerebrovascular Ultrasound
- Medical Imaging Techniques
Background:
- Power-based transcranial color-coded duplex sonography (p-TCCD) offers potential advantages over frequency-based TCCD (f-TCCD), including angle independence and reduced aliasing.
- p-TCCD aims to improve signal-to-noise ratio for enhanced ultrasonic imaging.
Purpose of the Study:
- To assess p-TCCD's capability in visualizing cerebral parenchyma blood flow.
- To compare the advantages, limitations, and velocity measurement reliability of p-TCCD against f-TCCD in basal cerebral arteries of healthy individuals.
Main Methods:
- Two investigators conducted 15 p-TCCD and 15 f-TCCD studies on 30 normal subjects.
- Peak systolic (Vs) and end-diastolic (Vd) velocities were measured in major cerebral arteries.
- Reliability was assessed using correlation coefficients (r) and coefficients of variation (CV) between p-TCCD and f-TCCD.
Main Results:
- p-TCCD successfully depicted arteries with a perpendicular course to the ultrasound beam but failed to visualize cerebral parenchyma flow.
- Velocity measurements demonstrated good reliability for p-TCCD (Vs: r=0.84-0.93, CV=7.7%-10.8%; Vd: r=0.87-0.90, CV=10.3%-13.7%).
- Limitations of p-TCCD, such as lack of directional information and tissue motion artifacts, were deemed unimportant.
Conclusions:
- In normal adults with adequate acoustic windows, p-TCCD demonstrated reliable velocity measurements in basal cerebral arteries.
- However, p-TCCD offered no significant advantages over f-TCCD and presented several minor limitations.