[A study on duration of using VCu200 intrauterine device]
Objective:
To make certain of the duration of using V-copper 200mm2 intrauterine device (VCu200 IUD).
Methods:
Six hundreds and sixty six pieces of VCu200 IUD used for different time period and removed for various reasons were analyzed. The life span of VCu200 IUD depended on corrosion, fragmentation and damage of copper wire and these parameters were measured on VCu200 IUD removed.
Results:
There were significant difference in copper wire fragmentation rate between tenth year-group and fifteenth year group which were 12.36%, 25.53% respectively (P < 0.001), and so were the expulsion rate of copper wire 2.54%, 9.57% respectively (P < 0.05). The copper loss and copper releasing rate in tenth year-group were 39.73%, 22.95%micrograms/day respectively. Copper corrosion resulted in fragmentation of copper wire which occurred the earliest in three years of use and expulsion of copper wire segment occurred after five years of use.
Conclusion:
Comparing the above results with other copper intrauterine devices, it was suggested that the suitable duration of VCu200 IUD use was ten years.
More Related Videos
05:51Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
08:06The Immediate Partial Removal of Cumulus-Oocyte Complexes: A Refined Approach for Rapid Observation of In Vitro Fertilization
Published on: October 18, 2024
Related Concept Videos
Uterine Tubes
Birth Control Methods
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Pharmacokinetic–Pharmacodynamic Relationship: Duration of Dose-Effect Relationship
Intrauterine Drug Delivery Systems
Urodynamic Studies: Uroflowmetry