Lung abscess due to Pseudomonas cepacia

R H Poe, H R Marcus, G L Emerson

    The American Review of Respiratory Disease
    |May 1, 1977
    PubMed
    Summary

    Ultrasonic nebulizers can harbor Pseudomonas cepacia, leading to lung abscesses in patients with pneumonia. Healthcare providers should suspect nosocomial infections when this bacterium is identified.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    More tort reform is needed.

    Texas medicine·2000
    Same author

    Managing cough as a defense mechanism and as a symptom. A consensus panel report of the American College of Chest Physicians.

    Chest·1998
    Same author

    Pulmonary mycetoma due to Candida albicans: case report and review.

    Clinical infectious diseases : an official publication of the Infectious Diseases Society of America·1996
    Same author

    Tropical pulmonary eosinophilia masquerading as acute bronchial asthma.

    Respiration; international review of thoracic diseases·1996
    Same author

    Include more patients in the study.

    Chest·1995
    Same author

    Use of fiberoptic bronchoscopy in the diagnosis of bronchogenic carcinoma. A study in patients with idiopathic pleural effusions.

    Chest·1994

    Area of Science:

    • Medical Microbiology
    • Infectious Diseases
    • Pulmonary Medicine

    Background:

    • Pneumonia is a common respiratory infection requiring treatment.
    • Ultrasonic nebulization is a method for delivering respiratory medications.
    • Diabetes mellitus can complicate infections and their management.

    Observation:

    • A diabetic patient with pneumonia developed a lung abscess after ultrasonic nebulizer therapy.
    • The causative agent was identified as Pseudomonas cepacia.
    • The patient had direct exposure to the nebulizer's reservoir via a modified medication cup.

    Findings:

    • The ultrasonic nebulizer's reservoir served as the source of Pseudomonas cepacia infection.
    • Contaminated nebulizer equipment can lead to nosocomial pneumonia and abscess formation.
    • Pseudomonas cepacia transmission occurred due to a breach in the device's sterile design.

    Implications:

    • Healthcare facilities must ensure rigorous disinfection and maintenance of respiratory therapy equipment.
    • Modified medical devices pose a risk for pathogen transmission and healthcare-associated infections.
    • Prompt identification of the source is crucial for preventing further patient infections and outbreaks.