Spirometric versus Fick-derived oxygen consumption: which method is better?

D N Thrush1

  • 1Department of Anesthesiology, University of South Florida College of Medicine, Tampa, USA.

Critical Care Medicine
|January 1, 1996
PubMed
Summary

The Fick equation for measuring oxygen consumption (VO2) showed significant bias compared to spirometry in pigs. These findings suggest caution when using Fick-derived VO2 for critical care decisions.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Cardiac sympathetic blockade during spinal anesthesia involves both efferent and afferent pathways.

Anesthesiology·2000
Same author

Vagotonia and cardiac arrest during spinal anesthesia.

Anesthesiology·1999
Same author

Cardiopulmonary resuscitation: effect of CPAP on gas exchange during chest compressions.

Anesthesiology·1999
Same author

Biventricular failure after pericardial window.

Journal of cardiothoracic and vascular anesthesia·1998
Same author

Is epinephrine contraindicated during cardiopulmonary resuscitation?

Circulation·1997
Same author

Does significant arterial hypoxemia alter vital signs?

Journal of clinical anesthesia·1997

Area of Science:

  • Physiology
  • Critical Care Medicine

Background:

  • Oxygen consumption (VO2) is crucial for assessing critically ill patients.
  • The Fick equation is a common method for calculating VO2, relying on cardiac output and arterial-venous oxygen content difference.

Purpose of the Study:

  • To compare the accuracy of VO2 measurements using the Fick equation against a spirometric technique.
  • To evaluate the interchangeability of these two methods for VO2 determination.

Main Methods:

  • A prospective study was conducted in a university laboratory setting using nineteen adult pigs.
  • VO2 was calculated via the Fick equation using cardiac output (thermodilution) and oxygen content (galvanic fuel cell).
  • A spirometric technique measured VO2 by oxygen disappearance rate; VO2 was pharmacologically varied (dobutamine, labetalol).

Related Experiment Videos

Main Results:

  • A bias of 58 mL/min and a precision (SD of bias) of 35 mL/min were observed between Fick-derived and spirometrically measured VO2.
  • Fick-derived VO2 values were consistently higher than those measured spirometrically.
  • A strong correlation (r=0.90) was found, but a significant unexplained difference persisted.

Conclusions:

  • Despite efforts to minimize error, a notable discrepancy exists between Fick and spirometric VO2 measurements.
  • The Fick method and spirometry are not interchangeable for VO2 determination.