Cochlear implant device failure: diagnosis and management

P R Kileny1, L Z Meiteles, T A Zwolan

  • 1Department of Otolaryngology--Head and Neck Surgery, University of Michigan Medical Center, Ann Arbor, Michigan 48109-0312, USA.

The American Journal of Otology
|March 1, 1995
PubMed
Summary

Receiver-stimulator failure in cochlear implants is rare but occurs. A 3% failure rate was observed in 6 patients, with device failure occurring 6 months to 3.5 years post-implantation.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Auditory Event-Related Potentials Elicited From Cochlear Implant Recipients and Hearing Subjects.

American journal of audiology·2015
Same author

Balance System Function.

American journal of audiology·2015
Same author

Prenatal Iron Deficiency and Auditory Brainstem Responses at 3 and 10 Months: A Pilot Study.

The Hong Kong journal of paediatrics : the journal of Hong Kong Paediatric Society = Hsiang-kang erh k'o i hsueh hui hui k'an·2015
Same author

Cochlin isoforms and their interaction with CTL2 (SLC44A2) in the inner ear.

Journal of the Association for Research in Otolaryngology : JARO·2007
Same author

Absent semicircular canals in CHARGE syndrome: radiologic spectrum of findings.

AJNR. American journal of neuroradiology·2006
Same author

Pediatric cochlear implant patient performance as a function of age at implantation.

Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology·2004

Area of Science:

  • Otolaryngology
  • Biomedical Engineering
  • Neuroscience

Background:

  • Cochlear implant receiver-stimulator failures, though uncommon, are a concern as patient numbers increase.
  • This study investigates a cohort of patients experiencing complete and irreversible device failure.

Observation:

  • Six out of over 200 patients at the University of Michigan Medical Center experienced receiver-stimulator failure.
  • Failure occurred between 6 months and 3.5 years after initial cochlear implant surgery.

Findings:

  • A 3% failure rate was determined for the Cochlear Corporation multichannel implant receiver-stimulator.
  • Device failure was diagnosed using psychophysical, electrophysiologic, and averaged electrode voltage measurements.
  • Averaged electrode voltage measurements were particularly effective in assessing implant condition, differentiating from physiological changes.

Related Experiment Videos

Implications:

  • Understanding device failure rates is crucial for managing patient expectations and device longevity.
  • Successful explantation and reimplantation procedures were performed on all affected patients.
  • This data informs future cochlear implant design and patient management strategies.