Reliability and concurrent validity of the BROM II for measuring lumbar mobility
J Breum1, J Wiberg, J E Bolton
1Anglo-European College of Chiropractice, Bournemouth, England.
The BROM II device shows good reliability for measuring lumbar flexion and lateral flexion. Further studies are needed to confirm its accuracy for extension and rotation in symptomatic patients.
Area of Science:
- Biomechanical measurement
- Spinal mobility assessment
- Chiropractic research
Background:
- Accurate measurement of lumbar mobility is crucial for diagnosing and managing spinal conditions.
- Existing methods for assessing spinal mobility may have limitations in reliability or validity.
- The BROM II ( biệt danh cho thiết bị đo lường ) is a novel device for quantifying spinal movement.
Purpose of the Study:
- To evaluate the intra- and inter-examiner reliability of the BROM II for measuring lumbar mobility.
- To determine the concurrent validity of the BROM II against the established double inclinometer method.
- To assess the device's performance across sagittal, coronal, and transverse planes of motion.
Main Methods:
- A blind intra- and inter-examiner reliability study was conducted.
- Forty-seven asymptomatic chiropractic students participated in the study.
- Lumbar mobility was measured in degrees using both the BROM II and a double inclinometer.
Main Results:
- The BROM II demonstrated good intra- and inter-examiner reliability for lumbar flexion (ICC: 0.91, 0.77) and lateral flexion (ICC: 0.91, 0.85).
- Reliability for extension (ICC: 0.63, 0.35) and rotation (ICC: 0.57, 0.36) was less supported.
- Concurrent validity between the BROM II and double inclinometer showed partial support (ICC range: 0.27–0.75).
Conclusions:
- The BROM II is a reliable instrument for measuring lumbar sagittal (flexion) and coronal plane mobility.
- Its reliability in extension and rotation requires further investigation.
- Additional research on symptomatic populations is necessary before clinical recommendation.