Related Experiment Video
Updated: Aug 13, 2026

Corneal Confocal Microscopy: A Novel Non-invasive Technique to Quantify Small Fibre Pathology in Peripheral Neuropathies
Published on: January 4, 2011
Short- and long-term repeatability of Visioptic Alcon EyeMap (Visioptic EH-270) corneal topographer on normal human
Abstract:
The purpose of this study was to examine the short-term and long-term (6 months) repeatability of the Alcon EyeMap (Visioptic EH-270)a Computerized Corneal Topographer on normal eyes. Three measurements were taken on both eyes of 39 subjects with the corneal topographer. Both eyes of seven of these subjects were measured 6 months later in order to evaluate long-term repeatability. The data showed that short-term reliability of the instrument was acceptable with a mean standard deviation (SD) < or = 0.5 D up to 3 mm (6-mm optical zone) from the center of the normal cornea. Beyond that distance, SD's of repeated measures increased systematically, making the results less useful. Long-term repeatability of the instrument over a 6-month period was acceptable clinically, although statistically significant differences were noted.

