Related Experiment Videos
Biocompatible sample pretreatment for immunochemical techniques using micellar liquid chromatography for separation
U Lövgren1, M Johansson, K Kronkvist
1Department of Bioanalytical Chemistry, Astra Draco AB, Lund, Sweden.
Abstract:
Micellar liquid chromatography (MLC) using Tween 20 as surfactant was evaluated as a biocompatible sample pretreatment preceding immunoassay in order to obtain an increased selectivity of the assay and a simplification of the sample pretreatment procedure. Different stationary phases and chromatographic conditions were studied for the separation of budesonide and cortisol and some steroids known to interfere in immunoassay of these compounds. The separation was dependent on several parameters, for example, temperature, the concentration of Tween 20, pH and ionic strength of the mobile phase, and nature of the stationary phase. A precolumn venting system was used, which allowed for 140 direct injections of 25 microliters of human blood plasma, without loss of chromatographic performance. Results obtained from the coupling of MLC to an immunoassay for cortisol illustrates the selectivity which can be obtained, and that simplification of the sample pretreatment is possible using this technique.