Fetal pulse oximetry. A methodogical study

A Maesel1, L Mårtensson, S Gudmundsson

  • 1Department of Obstetrics and Gynecology, University of Lund, Malmö, Sweden.

Acta Obstetricia Et Gynecologica Scandinavica
|February 1, 1996
PubMed
Summary

Fetal pulse oximetry monitoring during labor is safe and practical, with successful sensor placement in 95% of cases. This method provides valuable fetal oxygen saturation (SpO2) data, though further research is needed to confirm its clinical utility.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

ISUOG Safety Committee updated recommendation on use of respirators by practitioners undertaking obstetric and gynecological ultrasound in context of SARS-CoV-2 Omicron variant of concern.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2022
Same author

ISUOG statement on the non-diagnostic use of ultrasound in pregnancy.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2021
Same author

ISUOG statement on the safe use of Doppler for fetal ultrasound examination in the first 13 + 6 weeks of pregnancy (updated).

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2021
Same author

Extent of absent end-diastolic flow in umbilical artery and outcome of pregnancy.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2020
Same author

Infant outcome after active management of early-onset fetal growth restriction with absent or reversed umbilical artery blood flow.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2020
Same author

Experiences and needs concerning health related information for newly arrived refugees in Sweden.

BMC public health·2020

Area of Science:

  • Obstetrics and Gynecology
  • Fetal Monitoring
  • Medical Devices

Background:

  • Fetal pulse oximetry is an emerging technology for monitoring fetal oxygenation during labor.
  • This study evaluated the practical application of the FS-10 Oxisensor (Nellcor).
  • Obstetricians and midwives were involved in sensor placement, with SpO2 readings not influencing patient management.

Purpose of the Study:

  • To assess the practical applicability and safety of fetal pulse oximetry during labor.
  • To determine the success rate of sensor placement and data acquisition.
  • To observe changes in fetal arterial oxygen saturation (SpO2) during different stages of labor.

Main Methods:

  • Methodological study involving 96 singleton pregnancies.
  • Fetal pulse oximetry sensor (FS-10 Oxisensor) placement and monitoring during labor.

Related Experiment Videos

  • Calculation of mean SpO2 levels at specific cervical dilatations and in the second stage; statistical analysis using Student's t-test.
  • Main Results:

    • Successful sensor placement in 91% (95%) of cases.
    • Minor, transient adverse events: one case of maternal pain, two newborns with temporary impression marks.
    • Obtainable SpO2 values for 69% of the recording time (mean 134 minutes); significant SpO2 decrease observed from the first to the second stage of labor.

    Conclusions:

    • Fetal pulse oximetry appears safe for both mother and child.
    • The method enables SpO2 data acquisition during approximately two-thirds of the monitoring period.
    • Further research is required to establish the clinical value of fetal pulse oximetry in labor management.